9-8 Care and Maintenance
9.3.3 Peripherals and Options Check
If the system is not configured with any module or peripheral, the corresponding items checking can
be skipped.
Table 9-7 Options, Peripherals and Accessories Check list
No.
Content
Method
1.
Color and B/W video
printer
Check if the output of video printer is normal.
2.
Graph / text printer
Check if the output of graph / text printer is normal.
3.
Foot switch
Check if the foot switch can implement the set functions
according to the program.
4.
DVD-R/W
Check if DVD can work normally (write, read and pop).
5.
Bar code scanner
Check if the scanner can work normally and the output content
is right.
6.
DICOM
Check if DICOM can work normally and send pictures and
other data to DICOM server.
RemarkPlease refer to 5.3 for details.
9.3.4 Mechanical Safety Inspection
Mechanical safety inspection is mainly used to check mechanical strength and mechanical
function of the key assembly of ultrasonic system. The mode of test evaluation mainly is: Perform
the evaluation by means of visual check and operating check, if the check result cannot pass, the
system is in abnormal status now. Stop using the system and adopt proper measures. The test flow
is as following:
Check handle
Check control panel And
turning mechanism
Check display and turning
mechanism
Check other mechanics
Check the connection of
casters
Check casters
Fig 9-4 Mechanical Safety Inspection Flow

Table of Contents

Other manuals for Mindray Z6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z6 and is the answer not in the manual?

Mindray Z6 Specifications

General IconGeneral
BrandMindray
ModelZ6
CategoryMedical Equipment
LanguageEnglish

Related product manuals