3-2 System Installation
3.1.2 Installation Condition
3.1.2.1 Space Requirements
Place the system with the necessary accessories at a proper position for convenient use.
1. Place the system in a room with good ventilation or having an air conditioning unit.
2. Leave at least 20cm clearance around the system to ensure effective cooling.
3. A combination lighting system in the room (dim/bright) is recommended.
4. Except the receptacle dedicated for the ultrasound system, at least 3-4 spare receptacles
on the wall are available for the other medical devices and peripheral devices.
5. Power outlet and place for any external peripheral are within 2 m of each other with
peripheral within 1 m of the unit to connect cables.
3.1.2.2 Network Environment
Both wireless and wired LAN are supported by this ultrasound system.
Data transmission is allowed between different departments or areas without network cable.
Network can be automatically connected after disconnection in case that the device is required to
be moved, wireless transmission task can be recovered after the network resumed to normal
condition. Confirm the network devices and network conditions before the installation.
1. General information: default gateway IP address, and the other routers related information.
2. DICOM application information: DICOM server name, DICOM port, channels, and IP
address.
3.1.2.3 Confirmation before Installation
Please confirm the following items before installation:
1. The video format used in the region or country where the system is installed.
2. The language used in the region or country where the system is installed.
3. The power voltage and frequency used in the region or country where the system is
installed.
4. Obstetric formulae and other measurement formulae used in the region or country where
the system is installed.
5. Other settings to be used in the region or country where the system is installed but different
from the factory settings.
6. The doctor’s habits when using the system.
Please confirm the items above prior to the installation training, and do the system settings
according to the universal setting of installed region or country.
3.2 Unpacking
Tools Required: None.

Table of Contents

Other manuals for Mindray Z6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z6 and is the answer not in the manual?

Mindray Z6 Specifications

General IconGeneral
BrandMindray
ModelZ6
CategoryMedical Equipment
LanguageEnglish

Related product manuals