5-4 Function and Performance Checking Method
port have been verified,
remains are:
VGA port
Connect VGA display
Network port
USB port
those displayed on the ultrasound system display, no
character and image loss, no color difference, no fluttering and
flicking.
Smooth communication in both network and connection.
USB port data storage/accessing are normal.
5.4 Function Checks
NOTE:
A complete function inspection is described here, do the checking according to the
actual system configuration.
5.4.1 Check Flow
Check each imaging
mode
Check the measurements
Check in cine mode
Check the probe
application
Image & video manage
Record and save the
exam