25
Calibration
2
1
By pushing key twice, return to the measurement mode.
displays after the span calibration is finished.
Push key to complete the calibration mode.
Step
9
End of calibration
Measurement mode
▲
The display becomes , and the set data is memorized
internally.
F
ENTER
FUNC.FUNC.
• The set value is registering temporary until fixing by pushing the F/Enter key on
display. The set value is not preserved when canceling the calibration on
the way.
• Please set the fine adjustment of ZERO and SPAN, the digital linearization, ZERO
set, and Tare weight cancellation after completing the calibration. If the calibration
is executed over again, those data is cleared.
light on when the load display exceeds over [(+Weighing capacity)+9D] or [+110 %
of weighing capacity. (Setting by CF-03 of C function)
light on when the load display falls below the [(- Weighing capacity) - 9D], falls bellow
the [+110 % of weighing capacity], or falls below [-20D].
(Setting by CF-03 of C function)
Error display in the measurement and after the calibration