ER-0
ZERO set error.
becomes blinking display when you execute ZERO set
with not enforceable condition. By pushing F/ENTER key, [ER-0] continues the
blinking display. By pushing F/ENTER key, the error condition is released and
returns to the condition before the error is displayed.
ER-1
This displays with blinking for about 2 seconds when you select the setteing of
C function, function and S function by the number that doesn't exist. Meanwhile,
lignts on. (However, the setting of C function is excluded.
ER-2
This displays with blinking for about 2 seconds when you set the calibration
by [Weighing capacity (DISP) < Mass of weight (LOAD)], and when the
compensation value is the value of weighing capacity or more, or ZERO or less
in adjusting it for digital linearization.
ER-3 A/D conversion error. Please inform Minebea.
ER-7
Tare weight cancellation error.
becomes blinking display when you
execute tare weight cancellation with not enforceable condition. By pushing [F/
ENTER] key, [ER-7] continues the blinking display. By pushing F/ENTER key,
the error condition is released and returns to the condition before the error is
displayed.
ER-8
This displays with blinking for about 2 seconds when you set the value outside
the setting range for C function, function and S function. Meanwhile,
lights on.
ER-E EEPROM writing error. Please inform Minebea.
ER-R EEPROM reading error. Please inform Minebea.
TE-L
This displays with blinking for about 2 seconds when the load cell I/O value in
calibration becomes - 2.5 mV/V or less that falls below the ZERO adjustment
range at minus side.
TE-H
This displays with blinking for about 2 seconds when the load cell I/O value in
calibration becomes 2.5 mV/V or more that exceeds over the range of ZERO
adjustment range at plus side.
SP-L
This displays with blinking for about 2 seconds when the load cell I/O value is
less than 0 mV/V that doesn't reach to SPAN adjustment range.
SP-H
This displays with blinking for about w seconds when the load cell I/O value is
more than 3.1 mV/V that exceeds over SPAN adjustment range.
In the simple calibration, this display keeps blinking the error even if the condition
returns to the measurement mode, and the measurement is not executed. It
released after calibrating again by the right range.
LN-L
This displays with blinking for about 2 seconds when the load value is lower than
ZERO and does not satisfy the condition of LNR-1 < LNR2 < LNR3 in registering
the load value compensation by digital linearization.
LN-H
This displays with blinking for about 2 seconds when the load value is higher
than the weighing capacity in registering the load value compensation by digital
linearization.
-
(When C function CF-15 [Power on ZERO operation is valid.)
This displays when the power supply is turned on in the condition that the load
cell output exceeds over ± 10 % of the weighing capacity.
By pushing [FUNC.] key in the display of [
], the display becomes the
load display compulsorily.
19. Error display