41
Setting of Function
6-8.Power on ZERO
The power on ZERO function makes the display to a zero if the changes of display is within
± 10 % of the maximum capacity when the power is turn on, or in the stable condition in the
operation of display turning on.
6-8-1. Operation of power on ZERO
[VALID] or [INVALID] of power on ZERO is selected by C function CF-15.
Setting range0,1
0[INVALID] 
1{VALID]
6-9.Tare weight cancellation
By pushing [F/ENTER] key after allocating this function to [F/ENTER], (Tare weight
cancellation) and
(Net value) in the condition display section is light on at the same
time as the load display value becoming a net amount display, and the display becomes zero.
There are following two procedures to clear a tare weight cancellation.
After allocating [Tare weight cancellation] to [F/ENTER] key, and push it when gross
value is zero.
After allocating [Tare weight cancellation clear] to [F/ENTER] ke
y, and push it.
After tare weight cancellation is cleared,
in the condition display section is
light on at the same time as the display becomes gross value.
Both [Tare weight cancellation] and [Tare weight cancellation clear] can be used by not only
allocating to [F/ENTER] key but also allocating to external control input.
6-9-1. Operating condition of tare weight cancellation
The operating condition of tare weight cancellation is selected by C function CF-16.
Setting range0 3
0Operation during steady and 0 < [Gross value] <= [Weighing
capacity]
1Continuous operation during 0 < [Gross value] <= [Weighing capacity]
2Operation during steady and [Gross value] <= [Weighing capacity]
3Continuous operation during [Gross value ] <= [Weighing capacity]
The ZERO compensation is not accepted in the following condition when the setting
of C function is [Power on ZERO effective] and power on ZERO is starting.
When the display is not a stable condition, it becomes as all lighting.
When the measurement value exceeds ± 10 % of the maximum capacity,
displays. At this time, by pushing [FUNC.] key, the display becomes a
load value compulsorily.
When the power on ZERO is executed, each data of tare weight cancellation, xed
tare weight cancellation, ZERO set and ZERO tracking are all cleared.
 
If ZERO set is executed, tare weight cancellation is cleared.