41
Setting of Function
6-8.Power on ZERO
The power on ZERO function makes the display to a zero if the changes of display is within
± 10 % of the maximum capacity when the power is turn on, or in the stable condition in the
operation of display turning on.
6-8-1. Operation of power on ZERO
[VALID] or [INVALID] of power on ZERO is selected by C function CF-15.
Setting range:0,1
0:[INVALID]
1:{VALID]
6-9.Tare weight cancellation
By pushing [F/ENTER] key after allocating this function to [F/ENTER], (Tare weight
cancellation) and
(Net value) in the condition display section is light on at the same
time as the load display value becoming a net amount display, and the display becomes zero.
There are following two procedures to clear a tare weight cancellation.
• After allocating [Tare weight cancellation] to [F/ENTER] key, and push it when gross
value is zero.
• After allocating [Tare weight cancellation clear] to [F/ENTER] ke
y, and push it.
After tare weight cancellation is cleared,
in the condition display section is
light on at the same time as the display becomes gross value.
Both [Tare weight cancellation] and [Tare weight cancellation clear] can be used by not only
allocating to [F/ENTER] key but also allocating to external control input.
6-9-1. Operating condition of tare weight cancellation
The operating condition of tare weight cancellation is selected by C function CF-16.
Setting range:0 ~ 3
0:Operation during steady and 0 < [Gross value] <= [Weighing
capacity]
1:Continuous operation during 0 < [Gross value] <= [Weighing capacity]
2:Operation during steady and [Gross value] <= [Weighing capacity]
3:Continuous operation during [Gross value ] <= [Weighing capacity]
• The ZERO compensation is not accepted in the following condition when the setting
of C function is [Power on ZERO effective] and power on ZERO is starting.
① When the display is not a stable condition, it becomes as all lighting.
② When the measurement value exceeds ± 10 % of the maximum capacity,
displays. At this time, by pushing [FUNC.] key, the display becomes a
load value compulsorily.
• When the power on ZERO is executed, each data of tare weight cancellation, fixed
tare weight cancellation, ZERO set and ZERO tracking are all cleared.
• If ZERO set is executed, tare weight cancellation is cleared.