Calibration
4-8. Digital linearization
After the calibration, the measurement error of some scale interval levels might be caused
between ZERO and SPAN (maximum capacity) by the inuence of the measuring section
etc.The digital linearization function executes the compensation of three points or less
except ZERO and SPAN, and to reduce the measurement error.
4-8-1. Setting procedure of digital linearization mode
Changeover to digital linearization mode
Step
1
Measurement mode
Move to the digital
linearization mode.
By pushing key from normal measurement mode,
displays.
4
By pushing key twice, display changes in order of
Display becomes , you can enter into the digital
linearization mode.
.
FUNC.FUNC.
CAL-ZCAL-Z
2
3
Push key at three times. The display changes as follows:
Display becomes    . By pressing key,
displays.
CAL-ZCAL-Z
ENTERENTER
FF