37
Setting of Function
Reference No. Item Function No. Set value Contents
6-11
Stability detection in the
setting.
CF-20
● 0 Invalid
1 Valid
6-12-1
Method of setting
gravity acceleration
compensation
CF-25
● 0 Set the district number
1 Set the numerical value of gravity acceleration
6-12-2
District number of using
place
CF-26
(Can be set in
CF25=0)
01~16 Unit : District
● 10
6-12-3
District number of
calibration place
CF-27
(Can be set in
CF25=0)
01~
16 Unit : District
● 10
6-12-4
Gravity acceleration value
setting of using place
CF-28
(Can be set in
CF25=1)
9.000~
9.999
● 9.797
Unit : m/s
2
6-12-5
Gravity acceleration value
setting of calibration place
CF-29
(Can be set in
CF25=1)
9.000~
9.999
● 9.797
Unit : m/s
2
6-13
Calibration data
(for reference)
Scale interval
CF-90 At the shipment from factory : 1
6-13
Calibration data
(for reference)
Weighing capacity
CF-9
1 At the shipment from factory : 10000
6-13
Calibration data
(for reference)
mass of weight
CF-92 At the shipment from factory : 10000
6-13
Calibration data
(for reference)
ZERO mV/V value
CF-93 At the shipment from factory : 0.0000
6-13
Calibration data
(for reference)
SPAN mV/V value
CF-94 At the shipment from factory : 0.5000
6-14
Stability detection time
width at calibration
CF-97
00~99
Unit : 0.1 s
00 : OFF of
stability detection at calibration.
● 05
6-15 Digital linearize clear CF-98
The data of the compensation by the digital
linearization is cleared.
6-16 C function clear CF-99 C function set content is returned default setting.
● : default