EasyManua.ls Logo

MR ETOS ED - Hot-spot calculation (optional)

MR ETOS ED
406 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Visualization
Maschinenfabrik Reinhausen GmbH 2020264 6385142/08 ENETOS
®
ED
9.26.1 Hot-spot calculation (optional)
You can set the parameters for the hot-spot calculation in the Hot-spot
menu. If you use the device in the version with a winding temperature sen-
sor, then you can also select whether the hot-spot is to be measured or cal-
culated.
Hot-spot measurement
Hot-spot measurement requires that you connect [Section 9.28, Page 282]
a sensor for recording the winding temperature to the device via an MR sen-
sor bus and link [Section 9.29, Page 292] the transmitted data points to
the function of the analog input (winding temperature 1...24).
Hot-spot calculation
The device can calculate the hot-spot in accordance with the thermal models
of the standards IEC60076-7 and IEEEStdC57.91. To enable this, you
must set the calculation parameters.
Figure191: Hot-spot calculation
If you use the optional function "Hot-spot calculation on 3 different windings
(W1, W2, W3)", you must set the parameters "IEC: Gradient", "IEEE: Gradi-
ent", "IEC: Hot-spot factor" and "Nominal current" for W1, W2 and W3 re-
spectively.
If you are using a cooling system, you also need to enter the calculation pa-
rameters for each cooling stage in the menu under "CS x hot-spot".

Table of Contents

Related product manuals