138
Adjustment Level Section 7-4
Function The entire input range is shifted by a fixed rate (one-point shift). If the input
shift value is set to
−1.0°C, the SP is controlled to a value obtained by
subtracting 1.0
°C from the actual temperature.
Communications
Settings
See Also
• Related Parameter
“Input type” (initial setting level) (page 143)
7-4-7 Upper Limit Temperature Input Shift Value (insh)
Lower Limit Temperature Input Shift Value (insl)
Condition: Refer to the following table for details.
These parameters allow the input range to be shifted both at the upper and
lower limits (two-point shift) (as opposed to the “temperature input shift”
parameter, where the entire input range is shifted by a fixed rate (one-point
shift)). A two-point shift enables more accurate offset of the input range
compared with a one-point shift when the input shift values at the upper and
lower limits differ.
Conditions of Use
Note Two-point shift can be used with platinum-resistance thermometers and
thermocouples for upgraded pulse output models and analog output models only.
Function This parameter sets input shift values for each of the upper and lower limits
(two-point shift) of the input range.
Communications
ch Variable type Address
1C1 000F
2C1 010F
Setting range Unit Default
FFFFF831H to 0000270FH (−199.9 to 999.9) °C or °F0.0
Setting Upper limit temperature input shift value/
Lower limit temperature input shift value
Additional
temperature input
shift value display
ON
Input type Infrared temperature sensor Platinum-resistance
thermometer, thermocouple
(See note.)
Temperature input
shift type
- Two-point shift
Setting ch Variable type Address
Upper limit temperature
input shift value
1 C1 0010
2 C1 0110
Lower limit temperature
input shift value
1 C1 0011
2 C1 0111