Auto Focus Multi Code Reader V430-series User’s Manual (Z392) 7-13
I/O Parameters
Target Pattern
Always OFF
The target pattern will remain OFF at all times unless overridden by an operational command.
ON Only When Not in the Read Cycle
The target pattern is always ON except during the read cycle. If the operational command
overrides this setting, the target pattern will remain on at all times.
ON Only When in the Read Cycle
The target pattern will remain OFF except during the read cycle. If the operational command
overrides this setting, the target pattern will remain on at all times.
Always ON
The target pattern is always ON.
Usage: Assists users in positioning and locating symbols in the center of the
reader’s field of view.
Definition: The user can control when the targeting system is ON or OFF, and can
save this condition for power-on.
Serial Cmd: <K750,green flash mode,target pattern status,green flash duration>
Default: Always ON
Options: 0 = Always OFF
1 = ON only when not in the read cycle
2 = ON only when in the read cycle
3 = Always ON

Table of Contents

Other manuals for Omron V430-F Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron V430-F Series and is the answer not in the manual?

Omron V430-F Series Specifications

General IconGeneral
BrandOmron
ModelV430-F Series
CategoryBarcode Reader
LanguageEnglish

Related product manuals