Auto Focus Multi Code Reader V430-series User’s Manual (Z392) 13-23
Utilities
Other Operational Serial Commands
The following serial utility commands can be entered from Terminal or a PLC:
Train and Optimize
Train Symbol
Format: <TRAIN>: Initiates a Train operation.
This command will put the product into a mode of operation that will cause it to “train” the
next symbol decoded. This mode of operation will remain active until either a symbol is
decoded or the call is made to disable the mode and revert back to normal operation.
Upon decoding a symbol, the image processing will save pertinent information regarding
the target symbol to allow higher readability for the similar symbols.
The data collected by the Train operation can be saved for a power-on condition, using the
<Z> command.
Un-Train Symbol
Format: <UNTRAIN>: Initiates an Un-Train operation.
This command will cause the product to discard any information acquired during a Train operation.
The Untrained state can be saved for a power-on condition, using the
<Z>
command.
Train Status
Format: <TRAIN?>: Responds with Train Status
This command will return the current status of the TRAIN operation.
Response: < TRAIN,0>: Default, not trained
<TRAIN,1>: Train operation in process
<TRAIN,2>: Symbol trained
Train Persistence
The train state and parameters persist the same as ordinary parameters. Examples:
A unit in the trained state has not been saved. A power cycle will remove any trained
state information and the unit will come up in its configured state.
A unit is saved in a training state. The unit will come from power up in the training state
and train the first symbol read.
A unit is saved in a trained state. The unit will come from power up in the trained state
and only read the trained symbol type.
A trained unit is issued a reset default command <Ard>. The unit will return to an
untrained state but if power is cycled will return to its saved state.

Table of Contents

Other manuals for Omron V430-F Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron V430-F Series and is the answer not in the manual?

Omron V430-F Series Specifications

General IconGeneral
BrandOmron
ModelV430-F Series
CategoryBarcode Reader
LanguageEnglish

Related product manuals