Auto Focus Multi Code Reader V430-series User’s Manual (Z392) 8-19
Symbol Quality
Fixed Pattern Damage
Axial Non-Uniformity
Definition: Fixed Pattern Damage is calculated in accordance with the ISO 15415
specification.
When enabled, the Fixed Pattern Damage is attached to the symbol
output data as a value.
Serial Cmd: <K725, aperture, overall, contrast, modulation, reflectance margin, fixed
pattern damage, axial non-uniformity, grid non-uniformity, unused
ECC>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Definition: Axial Non-Uniformity is calculated in accordance with the ISO 15415
specification. The Axial Non-Uniformity grading is defined as follows.
A (4.0) if AN
≤ 0.06
B (3.0) if AN
≤ 0.08
C (2.0) if AN
≤ 0.10
D (1.0) if AN
≤ 0.12
F (0.0) if AN
> 0.12
When enabled, the Axial Non-Uniformity is attached to the symbol out-
put data as a value.
Serial Cmd: <K725, aperture, overall, contrast, modulation, reflectance margin, fixed
pattern damage, axial non-uniformity, grid non-uniformity, unused
ECC>
Default: Disabled
Options: 0 = Disabled 1 = Enabled