EasyManua.ls Logo

POLYSONICS DCT-7088 - Page 144

POLYSONICS DCT-7088
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7088-8000
-144
Figure F-13 Configuration Screen 4
General Configuration Procedures
Follow these general steps when configuring the flowmeter:
1. Go to the
CONFIGURATION
tab.
2. Follow the step by step procedure in Sections F.4.1 through F.4.11 to set the
parameters for each option (for example,
PIPE
,
TRANSDUCER
, and
FLOW
).
NOTE 1: Measurement units (
English versus metric) may be selected in the Pipe OD Setup.
NOTE 2:
If a value of less than 1 is entered, it must be preceded by a 0 (zero). For
example, if entering a pipe wall thickness of .5 inches, the number must be
entered as 0.5.
NOTE 3:
If an option or a text box is dimmed out (greyed out), it is currently not available
or contains view-only information which cannot be edited for the current
parameters.
3. After entering all parameters, apply the settings to the flowmeter by pressing the
SEND
command button (Figure F.14).
Figure F-14 Receive and Send Command Buttons

Table of Contents