7088-8000
-55
To delete a data log file:
1. Access Menu 80.
2. Select
VIEW
.
3. Scroll through the data log files with the
UP
or
DOWN
ARROW
keys to display the file that is to
be deleted.
4. Press
ENTER
.
The following is displayed:
5. Press the
5
key.
The following is displayed:
6. Press the decimal point (
.
) key.
The following is displayed:
Data Log Interval (Menu 81)
Menu 81 is used to set the interval for the data log.
This menu can be accessed directly (by pressing
MENU
and 81) or by selecting
INTERVAL
in Menu 80.
The data log interval must be entered in whole seconds
with a minimum interval of 1 second.
3.6.13 DIAGNOSTICS DISPLAY MENUS
The
DIAGNOSTICS
sub-menu contains various view-
only menus. These menus display important diagnostic
parameters that are currently used or calculated by the
flowmeter. These parameters are helpful when
configuring or troubleshooting the instrument.
Signal Strength/Margin (Menu 90)
This menu displays the signal strength and the margin.
The signal strength value displayed in this menu is the
average of the signal strengths for the upstream and
downstream transducers. Margin is an indicator of
signal quality. Margin is generally greater than 5% and
signal strength is generally greater than 3% under good
measurement conditions.
Erase Log File?
Are you sure?(5=Yes)
Erase Log File?
Really sure?(.=Yes)
Data log
Erased
Log Interval
60 secs
SigStr = 0%
Margin = 0%