7088-8000
-30
3.6 Primary Displays, Setup Menus, and Diagnostic Menus
This section describes the functions of the Primary Displays, the setup menus, and the
diagnostic menus. These displays and menus are described in numerical order according to
their menu addresses. The menus are also are listed under their appropriate sub-menus (for
example
PIPE
or
LINER
). These displays and menus can be accessed using cursor scrolling
(Section 3.3) or direct access (Section 3.2).
The Primary Displays are for viewing only and are not configurable. The setup menus are
used primarily to enter configuration data or to view the flowmeter’s current configuration
settings. Diagnostic menus are used for viewing various diagnostic parameters and are not
configurable. In addition, certain data may be printed by entering one of three available
print commands (Section 3.6.14).
IMPORTANT:
The flowmeter ships from the factory with the system password and scale
factor password disabled. To maximize security, new passwords should be
entered immediately. Refer to Menus 47 and 48 in Section 3.6.8.
3.6.1 PRIMARY DISPLAYS
This section describes the Primary Displays which
display the values for the flow rate, totalizers, velocity,
signal strength, or low signal cutoff (Menu 00
through 04). These displays are for viewing only.
(They are not configurable.)
NOTE:
Refer to Section 3.6.7 for information on
enabling and configuring the totalizers.
Flow/Net Totalizer (Menu 00)
The Flow/Net Totalizer display is the standard display
used under normal operating conditions. It displays the
flow rate and the net totalizer value. If the net totalizer
is not currently enabled, this display indicates the last
net totalized value.
Flow/Velocity (Menu 01)
The Flow/Velocity display indicates the flow rate and
fluid velocity. Velocity is displayed in feet per second
(FPS) if
ENGLISH
is selected as the measurement unit
in Menu 42 and in meters per second if
METRIC
is
selected.
Flow= 0.00 GPM
Net 0 x0.1Gal
Flow= 0.00 GPM
Vel = 0.00 FPS