7088-8000
-171
3. Click on the
UNIT ID
text box
4. Enter the identification number of the unit, using any whole number between 1 and
60,000.
F.4.11 DISPLAY OPTION GROUP
The
DISPLAY
option group is accessed by clicking on
DISPLAY
in the Options Column.
This function is applicable only to DCT1088 flowmeters which have the optional display.
The display can be configured to indicate any combination of flow rate, totalizer, or signal
strength values. For multiple selections, a cycle time can be entered to allow the display to
alternate between selections.
To configure the display:
1. Click on one or more of the following check boxes in the
DISPLAY
options group:
!
FLOW RATE
!
SIGNAL STRENGTH
!
TOTALIZER
2. If more than one check box was selected, click on the
CYCLE TIME
text box and enter
the cycle time (in whole seconds).
3. If
TOTALIZER
was checked, click on one of the following radio buttons (one only):
!
NET
(net totalizer)
!
POS
(positive totalizer)
!
NEG
(negative totalizer)
NOTE 1:
The following flow units cannot be represented on the display of the flowmeter:
!
Cubic feet
!
Oil barrels
!
Liquor barrels
!
Feet
!
Meters
If one of these flow or totalizer units is selected, a blank space is left where the
unit would normally be located on the display. The flow or totalizer numeric
value, however, is still displayed regardless of which unit is selected.
NOTE 2:
The totalizer reading on the flowmeter display is followed by a slash symbol
/
(a
“per” symbol) if a totalizer multiplier other than 1 is selected. For example, a
totalizer value of 2500 cubic meters with a totalizer multiplier of
x100
would be
displayed as:
25 M3/
.