7088-8000
-29
4. Complete the configuration process by accessing one of the Primary Displays
(Menu 00 through 04).
IMPORTANT:
The flowmeter will not use the new parameters until one of the Primary
Displays is accessed.
3.5 Quick Setup
The Quick Setup procedure provides the minimal steps required for configuring the
flowmeter to enable it to calculate transducer spacing, acquire ultrasonic signal, and
measure flow. The Quick Setup procedure is also provided on the Quick Setup Card.
The Quick Setup procedure is as follows:
1. Select a proper transducer site, following the guidelines in Section 4.1.
2. Enter the pipe OD (Menu 10).
3. Enter the pipe ID (Menu 12).
4. Enter the pipe material (Menu 13).
5. Enter the liner material if any (Menu 16).
6. Enter the liner thickness if any (Menu 17).
7. Enter the fluid type (Menu 20).
8. Enter the transducer type (Menu 23).
9. Enter the transducer mounting method (Menu 24).
Once these critical setup parameters are entered, the flowmeter calculates the proper
transducer spacing. The spacing is displayed in Menu 25.
10. Enter the flow rate units (Menu 30).
11. Install the transducers on the pipe, using the spacing measurement provided by the
flowmeter (Section 4.2) and connect the transducer cables to the flowmeter.
12. Access one of the Primary Displays (Menu 00 through 04) to complete the
configuration process.
The flowmeter is now capable of measuring velocity and flow.
IMPORTANT:
If any of the above setup parameters are changed, the flowmeter stops
measuring flow until the new value is entered and until one of the Primary
Displays (Menu 00 through 04) is accessed to accept the new value.