CubiScan 110 27
Chapter 4
Calibration
This chapter provides instructions for calibrating the CubiScan 110.
The CubiScan 110 is calibrated at the factory; however, some cir-
cumstances in which recalibration may be required include the fol-
lowing:
• Calibrate the CubiScan 110 if you have problems cubing and
weighing after assembly and setup.
• Calibrate the CubiScan if it is subjected to any type of mechani-
cal shock or collision with a heavy object.
• Calibrate the CubiScan as part of a regular maintenance sched-
ule. If the CubiScan is used heavily, scale calibration should be
performed monthly and sensor calibration yearly.
NOTE
If an error message appears during calibration, power the Cubi-
Scan off and back on and start calibration over (refer to “Computer
Error Messages” on page 45 or “Control Panel Error Messages”
on page 46 for more information).
Before You Begin
Before calibrating the CubiScan 110, remove all packages or other
material from the platform, and blow any dust off the sensor
screens. Refer to page 39 for information on cleaning the sensors.
☞