Calibrating Using the Control Panel Calibration
36 CubiScan 110
Calibrating the Scale
To calibrate the CubiScan scale, you need an official test weight in
the range of 50–100 pounds (25–45 kg) (it is recommended that you
calibrate with the maximum weight).
IMPORTANT
Do not begin scale calibration until you have the test weight. Cali-
brating without an accurate known weight (within .01 of a lb/kg)
can make all future weight readings inaccurate.
Take the following steps to calibrate the CubiScan scale.
NOTE
When calibrating the scale, the CubiScan must be stable with no
movement of the platform such as that caused by vibration or air
movement.
1. If you calibrated the sensors first, turn the CubiScan off and on,
then wait about fifteen seconds for self-diagnostics to finish
before calibrating the scale.
2. Make certain that the CubiScan platform is free of all objects.
3. Press on the control panel. After a short processing delay,
the WEIGHT window displays
0.00, and the DIM-FACTOR
window displays a command code of
11.
4. Make sure there is nothing on the CubiScan platform, and press
<msr>.
Do not disturb the CubiScan until the WEIGHT window dis-
plays the calibration weight (the weight of the last test weight
used to calibrate the scale).
5. If the calibration weight displayed matches the test weight you
are using, proceed to the next step.
lb
kg
------

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantronix CubiScan 110 and is the answer not in the manual?

Quantronix CubiScan 110 Specifications

General IconGeneral
BrandQuantronix
ModelCubiScan 110
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals