CubiScan 110 Command Set Communications Protocol
68 CubiScan 110
Zero
This command should be issued periodically to force the CubiScan
to perform internal compensations to adjust to changes in tempera-
ture and humidity. This command should only be issued when the
measuring field is free of packages! Never issue this command
when a package is present in the measuring field.
Pos Len Description Type Range ASCII
Command Format
1 1 Start of Text Control (STX) 02h
2 1 Zero Command Alpha (Z) 5Ah
3 1 End of Text Control (ETX) 03h
4 1 Carriage Return Control (CR) 0Dh
5 1 Line Feed Control (LF) 0Ah
Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Zero Command Alpha (Z) 5Ah
3 1 Acknowledge Alpha (A) 41h
4 1 End of Text Control (ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
6 1 Line Feed Control (LF) 0Ah
Negative Acknowledge Format
1 1 Start of Text Control (STX) 02h
2 1 Zero Command Alpha (Z) 5Ah
3 1 Neg. Acknowledge Alpha (N) 4Eh
4 1 End of Text Control (ETX) 03h
5 1 Carriage Return Control (CR) 0Dh
6 1 Line Feed Control (LF) 0Ah