Calibration Calibrating Using the Control Panel
CubiScan 110 37
If not, change the displayed weight as follows:
First, locate the cursor ( _ ) in the DIM-WEIGHT window
(under one of the calibration weight digits). The cursor position
determines which digit can be changed. If necessary, move the
cursor to a different digit by pressing to move to the left
or to move to the right. To increase the amount of the digit
at the cursor, press ; to decrease the digit, press
<zero>.
6. Place the test weight in the center of the CubiScan platform
(exact positioning is not required). Allow the scale to com-
pletely stabilize.
7. Press
<msr>. The test weight value is displayed in the
WEIGHT window, and the CubiScan calibrates the scale based
on the test weight. This completes scale calibration.
If you need to calibrate the sensors, refer to page 31 for instructions.
To prepare the scale for operation, refer to the section below for
instructions.
Preparing for Operation
To return the CubiScan to operation mode after calibration, proceed
as follows.
1. Turn off the power switch (on the power strip), and disconnect
the power cord from the power strip.
2. Carefully pull the controller box forward out of the base until
you can remove the cover.
3. Slide the controller box cover off to expose the DIP switch.
lb
kg
------
dom
intl
------------
in
cm
--------