- 115 -
format
FIS_I <CR><LF>
- command understood but not accessible at this moment
n - parameter, decimal value determining filter number
n 1 very fast
2 – fast
3 average
4 slow
5 very slow
CAUTION:
The numbering is assigned to a particular filter name and it is identical for all moisture
analyzer types.
The command changes settings for a current working mode if, for a particular moisture
analyzer type, filter settings are assigned to the working mode.
Example:
Command:
FIS_3<CR><LF>
- set average filter
Response:
FIS_OK<CR><LF>
- average filter set
Give current filter setting
Command overview:
Command gives information about currently set filter.
Format: FIG <CR><LF>
Response options:
FIG_x_OK<CR><LF>
- command carried out, response: currently set filter
FIG_I <CR><LF>
- command understood but not accessible at this
moment
x - parameter, symbol of currently set filter
Example:
Command:
FIG<CR><LF>
- give current filter
Response:
FIG_2_OK<CR><LF>
- currently set filter: average
Set value release
Format: ARS_n <CR><LF>
Response options:
ARS_OK <CR><LF>
- command carried out
ARS_E <CR><LF>
- error in-course of command execution, no parameter or incorrect
format
ARS_I <CR><LF>
- command understood but not accessible at this moment
n - parameter, decimal value determining value release options
n 1 – fast
2 fast + reliable
3 reliable
CAUTION:
The numbering is assigned to a particular value release option and it is identical for all
moisture analyzer types.
If for a particular moisture analyzer type, last digit settings are assigned to a working mode,
the command changes the current mode's settings.
Example:
Command:
ARS_2<CR><LF>
- set value release parameter to fast+reliable
option