Instrument Functions
R&S
®
NRP2
109User Manual 1173.9157.02 ─ 03
"Normal"
Displays the frequencies of the primary and, if used, secondary chan-
nel.
The following information is additionally displayed to single-channel
power measurement results in time slots and time gates (with R&S
NRP-Z78x wideband sensors only):
"Pk, Pk/Av" main value Av
"Av, Pk/Av" main value Pk
"Av, Pk" main value Pk/Av
"Extremes"
Displays the maximum, the minimum and the max-min values since the
search for extreme values has been started.
"Statistics"
Displays the longterm mean, the standard deviation and the total num-
ber of measurement results which have been evaluated since the
search for statistic values has been started.
SCPI command:
CALCulate<[1]..8>:DATA? on page 223
CALCulate<[1]..8>:MAXimum:DATA? on page 230
CALCulate<[1]..8>:MINimum:DATA? on page 230
CALCulate<[1]..8>:PTPeak:DATA? on page 230
CALCulate<[1]..8>:MEAN:DATA? on page 230
CALCulate<[1]..8>:SDEViation:DATA? on page 232
CALCulate<[1]..8>:COUNt:DATA? on page 223
DISPlay[:WINDow]<[1]..4>:AVALue on page 239
4.4.4 Relative measurement
The "Relative" function allows you to relate a power or a power ratio to a reference value.
The reference value can be a previously stored reading or an entered value. The
R&S NRP2 calculates the ratio of measurement result and reference value and the rel-
ative deviation of the measured value from the reference value and indicates the result.
Configuring measurements