1GP71_0E 166 Rohde & Schwarz
Status Command Function / Remark / Limitation
DCL Device clear.
GTL Goto local mode.
LLO Local lockout.
REN Goto remote mode.
SDC
Selective device clear.
Remark:
SDC applies an IP command to the emulation software and
presets the instrument.
Table 159: RC3102 GPIB functions
Commands
The following table shows the commands, supported by the RC3102.
Additional
Status Command / Function / Remark / Limitation
EX
Return to SCPI mode.
Remark:
This command has been added to the RC3102 command set.
The emulation parameters (content of registers 0 99) are NOT persistent and get
lost when leaving the emulation mode.
Table 160: Additional commands
Frequency
Status Command / Function / Remark / Limitation
FQ [value]
Select fixed frequency function and fixed frequency mode, optionally set fixed
frequency.
Value:
value in HZ | KZ | MZ | GZ
Default:
100.0 MZ
FR value
Set frequency offset.
Value:
value in HZ | KZ | MZ | GZ
Default:
0.0 MZ