HP8340 emulation mode
1GP71_0E 65 Rohde & Schwarz
Instrument State / Registers
Status Command / Function / Remark / Limitation
IP
Preset instrument.
Remarks:
HP8340 presets to the last saved backup register set 0.
SM* always presets to the HP8340 factory defaults and NOT to the last saved
backup register set 0.
SV<value>
Save the actual instrument state to the backup register.
Value:
1 9
Default:
1
Remarks:
The instrument state is NOT persistent and is lost when the instrument is switched
off or emulation mode is left.
RC value
Recall an instrument state from the backup register.
Value:
1 9
Default:
1
Remarks:
The instrument state is NOT persistent and is lost when the instrument is switched
off or emulation mode is left.
SHSV
Lock access to backup registers.
SHRC
Unlock access to backup registers.
AL1[value]
Turn alternate sweep mode on, optionally selects background (backup) register
<value>.
AL0
Turn alternate sweep mode off.
Table 57: Instrument State / Registers commands
Power Level and Control
Status Command / Function / Remark / Limitation
PL [value]
Select and optionally set output power level.
Value:
powermin … powermax [DB]
Default: