RC3102 emulation mode
1GP71_0E 168 Rohde & Schwarz
• Modulation
Status Command / Function / Remark / Limitation
AM value
Select amplitude modulation function, optionally set modulation depth.
Value:
value in PCT | %
Default:
0.0 PCT
Remark:
% is not supported.
MA0
Turn amplitude modulation off.
MA1
Turn amplitude modulation on.
Remark:
AM can be combined with FM and HM, but NOT with PM.
MA2
Select internal 400 Hz sine wave as modulations source.
MA3
Select internal 1000 Hz sine wave as modulations source.
MA4
Select external modulation source in AC coupling mode.
MA5
Select external modulation source in DC coupling mode.
PM
Select pulse modulation function.
MP0
Turn pulse modulation off.
MP1
Turn pulse modulation on.
Remark:
PM can be combined with FM and HM, but NOT with AM.
MP2
Select internal 400 Hz sine wave as modulations source.
MP3
Select internal 1000 Hz sine wave as modulations source.
MP4
Select external modulation source in AC coupling mode.
MP5
Select external modulation source in DC coupling mode.
FM value
Select frequency modulation function, optionally set modulation deviation.
Value:

Table of Contents