HP8360 emulation mode
1GP71_0E 88 Rohde & Schwarz
Status Command / Function / Remark / Limitation
MODulation:STATe?
Query the general (activation) state of the modulation.
Table 80: Modulation subsystem
POWer Subsystem
Status Command / Function / Remark / Limitation
POWer{:ALC:BANDwidth|:BWIDTh} value
POWer{:ALC:BANDwidth|:BWIDTh}:AUTO value
POWer:ALC:CFACtor value
POWer:ALC:SOURce value
Set / Query the type of leveling for output power automatic leveling
Value:
INTernal | DIODe | PMETer | MMHead
Default:
INTernal
Remark:
MMHead is not supported.
POWer:ALC[:STATe] value
Set / Query the (activation) state of leveling for output power automatic leveling
Value:
ON | OFF | 1 | 0
Default:
OFF
POWer:AMPLifier:STATe
POWer:AMPLifier:STATe:AUTO
POWer:ATTenuation value
Set / Query the output attenuation level.
Value:
num [DB] | UP | DOWN | MAXimum | MINimum
0.0 dB _ num _ 110.0 dB
Default:
---
Remark:
UP, DOWN, MAXimum, MINimum are not supported.
POWer:ATTenuation:AUTO value
POWer:CENTer value