HP8360 emulation mode
1GP71_0E 80 Rohde & Schwarz
Status Command / Function / Remark / Limitation
*TRG
Execute a trigger function.
*TST?
Performs a self test and returns result.
Remark:
0 is return in any case, indicating no error.
*WAI
Wait until pending operations have taken place.
Table 68: Common commands
ABORt Subsystem
Status Command / Function / Remark / Limitation
ABORt
Abort and reset sweep in process.
Table 69: Common commands
AM Subsystem
Status Command / Function / Remark / Limitation
AM[:DEPTh] value
Set / Query the modulation depth.
Value:
num [PCT] | MAXimum | MINimum | num [DB]
0.0 % _ num _ 100.0 %
Default:
50.0 %
Remark:
MAXimum, MINimum is not supported.
num DB is not supported.
AM:INTernal:FREQuency value
Set / Query the modulation rate.
Value:
num [HZ | KHZ | MHZ | GHZ] | MAXimum | MINimum
Fmin _ num _ Fmax
Default:
1.0 KHz
Remark:
MAXimum, MINimum is not supported.
AM:INTernal:FUNCtion value
AM:MODE value