AN680XX emulation mode
1GP71_0E 47 Rohde & Schwarz
Status Command / Function / Remark / Limitation
[:SOURce:]MARKer<n>:FREQuency value
Set / Query frequency of the selected marker.
Value:
freqmin … freqmax [HZ | KHZ | MHZ | GHZ] | MAX
Default:
M1 = 2.0 GHz, M2 = 8.4 GHz, M3 = 2.0 GHz, M4 = 5.0 GHz, M5 = 8.4 GHz
M6 = 8.4 GHz, M7 = 8.4 GHz, M8 = 8.4 GHz, M9 = 8.4 GHz, M10 = 3.5 GHz
[:SOURce:]MARKer<n>:POLarity value
Set / Query pulse polarity of markers.
Value:
NEGative | POSitive
Default:
POSitive
[:SOURce:]MARKer<n>:STATe value
Set / Query the state of the specified marker.
Value:
ON | OFF | 1 | 0
Default:
All OFF
[:SOURce:]MARKer<n>:VIDeo value
[:SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] value
Set / Query the power level of the RF output.
Value:
powermin … powermax [DBM] | MIN | MAX | UP | DOWN
Default:
0.0 dBm
[:SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude]:STEP[:INCRement] value
Set / Query the power level increment step size.
Value:
powermin … powermax [DB]
Default:
0.1 dB
[:SOURce:]POWer[:LEVel]:ALTernate value
[:SOURce:]POWer:ALC:SOURce value
[:SOURce:]POWer:ATTenuation value
Set / Query the output attenuator
Value:
0 110 [DB]
Default:
0.0 dB