2 Command System RIGOL
DS2000E Programming Guide
2-85
:CURSor:MANual:SOURce
Syntax
:CURSor:MANual:SOURce <source>
:CURSor:MANual:SOURce?
Description
Set or query the channel source of manual measurement with cursor.
Parameter
Name Type Range Default
<source> Discrete {CHANnel1|CHANnel2|MATH|NONE} CHANnel1
Note: Only channels currently enabled can be selected as the channel source.
Return Format
The query returns CHAN1, CHAN2, MATH, or NONE.
:CURSor:MANual:TUNit
Syntax
:CURSor:MANual:TUNit <unit>
:CURSor:MANual:TUNit?
Description
Set or query the horizontal unit in manual measurement with cursor.
Parameter
Name Type Range Default
<unit> Discrete {SECond|HZ|PERCent|DEGRee} SECond
Explanation
SECond: in the measurement results, A->X, B->X, and X are in "s" and 1/X is in "Hz".
HZ: in the measurement results, A->X, B->X, and X are in "Hz" and 1/X is in "s".
PERCent: in the measurement results, A->X, B->X, and X are expressed in percentage.
DEGRee: in the measurement results, A->X, B->X, and X are in ".
Return Format
The query returns SEC, HZ, PERC, or DEGR.