RIGOL 2 Command System
2-324 DS2000E Programming Guide
:WAVeform:XINCrement?
Syntax
:WAVeform:XINCrement?
Description
Query the time difference between two neighboring points of the channel source currently selected (refer
to the
:WAVeform:SOURce command) in the X direction.
Explanation
The return value is related to the current data reading mode:
In NORMal mode, XINCrement = TimeScale/100.
In RAW mode, XINCrement = 1/SampleRate.
In MAXimum mode, XINCrement = TimeScale/100 when the instrument is in running status;
XINCrement = 1/SampleRate when the instrument is in stop status.
The unit is related to the current channel source:
When the channel source is CHANnel1 or CHANnel2, the unit is s.
When the channel source is MATH with FFT operation or the channel source is FFT, the unit is Hz.
Return Format
The query returns the time difference between two neighboring points of the channel source currently
selected in the X direction in scientific notation.