2 Command System RIGOL
DS2000E Programming Guide
2-139
:MEASure:ADISplay
Syntax
:MEASure:ADISplay <bool>
:MEASure:ADISplay?
Description
Enable or disable all measurement; or query the current status of all measurement.
Parameter
Name Type Range Default
<bool> Bool {{1|ON}|{0|OFF}} 0|OFF
Explanation
All measurement can measure all the time and voltage parameters of the current measurement
source. Each measurement source has 21 measurement parameters and you can measure the three
measurement sources (CH1, CH2, and MATH) at the same time.
11 voltage measurement items: maximum, minimum, peak-peak, top, bottom, amplitude, average,
RMS, period RMS, overshoot, and preshoot.
8 time measurement items: period, frequency, rise time, fall time, positive pulse width, negative
pulse width, positive duty cycle, and negative duty cycle.
2 other measurement items: area and period area.
Return Format
The query returns 1 or 0.
:MEASure:AMSource
Syntax
:MEASure:AMSource <chan>[,<chan>[,<chan>]]
:MEASure:AMSource?
Description
Set or query the signal source of the all measurement parameters.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} CHANnel1
Return Format
The query returns one or more of CHAN1, CHAN2, and MATH. Multiple signal sources are separated by
commas.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals