:SAVE:CSV[:STARt]
Syntax
:SAVE:CSV[:STARt] <file_spec>
Description
Save the waveform data on the screen into the external memory in "*.csv" format.
Parameter
Name Type Range Default
<file_spec> ASCII string Refer to Explanation --
Explanation
<file_spec> includes the file storage directory and the filename with suffix. Its length cannot exceed
64 characters. Wherein, the storage directory can only be D:\. If the specified directory already
contains a file with the same filename, the original file will be overwritten.
The "*.csv" file stored includes the current horizontal time base, start point time and voltage value
corresponding to each point of the channel turned on as shown in the figure below. You can open
and edit the file via Excel.
Number of the point Start point time Horizontal time base
Voltage value corresponding to each point
Example
:SAVE:CSV D:\123.csv /*Store the waveform data on the screen into the external memory with the
filename 123.csv*/

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals