RIGOL 2 Command System
2-240 DS2000E Programming Guide
:TIMebase[:MAIN]:OFFSet
Syntax
:TIMebase[:MAIN]:OFFSet <offset>
:TIMebase[:MAIN]:OFFSet?
Description
Set or query the horizontal offset of the main time base and the unit is s.
Parameter
Name Type Range Default
<offset> Real
The range of <offset> is related to the horizontal time base mode (refer
to the
:TIMebase:MODE) and the operating state of the oscilloscope.
Y-T mode
RUN: When TimeScale is less than 10ms/div, the range of <offset>
is from (-MemDepth/SampleRate) to 1s.
When TimeScale is greater than or equal to 10ms/div, the
range of <offset> is from (-MemDepth/SampleRate) to (100 x
TimeScale).
STOP: -7000s to 7000s
Roll mode
RUN: This command is not available.
STOP: -7000s to 0
0
Note:
For the TimeScale, refer to the
:TIMebase[:MAIN]:SCALe command.
For the MemDepth, refer to the
:ACQuire:MDEPth command.
For the SampleRate, refer to the
:ACQuire:SRATe? command.
Return Format
The query returns the horizontal offset of the main time base in the scientific notation.