RIGOL 2 Command System
2-172 DS2000E Programming Guide
:MEASure:PWIDth
Syntax 1
:MEASure:PWIDth <chan>
Description 1
Enable the positive pulse width measurement function of the specified channel.
Syntax 2
:MEASure:PWIDth? <chan>
Description 2
Enable the positive pulse width measurement function of the specified channel and return the
measurement result.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the positive pulse width measurement result in scientific notation. When the
measurement result is invalid, the query returns 9.9e37.
Example
:MEASure:PWIDth CHANnel1
:MEASure:PWIDth? CHANnel1
You can use the following commands to enable the positive pulse width measurement function of the
specified channel and the statistic function of the measurement results as well as query the statistic
maximum value, minimum value, current value, average value, and standard deviation of the positive
pulse width measurement results.
:MEASure:PWIDth:SMAXimum
:MEASure:PWIDth:SMINimum
:MEASure:PWIDth:SCURrent
:MEASure:PWIDth:SAVerage
:MEASure:PWIDth:SDEViation