RIGOL 2 Command System
2-174 DS2000E Programming Guide
:MEASure:RTIMe
Syntax 1
:MEASure:RTIMe <chan>
Description 1
Enable the rise time measurement function of the specified channel.
Syntax 2
:MEASure:RTIMe? <chan>
Description 2
Enable the rise time measurement function of the specified channel and return the measurement result
(the unit is s).
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the rise time measurement result in scientific notation. When the measurement result
is invalid, the query returns 9.9e37.
Example
:MEASure:RTIMe CHANnel1
:MEASure:RTIMe? CHANnel1
You can use the following commands to enable the rise time measurement function of the specified
channel and the statistic function of the measurement results as well as query the statistic maximum
value, minimum value, current value, average value, and standard deviation of the rise time
measurement results.
:MEASure:RTIMe:SMAXimum
:MEASure:RTIMe:SMINimum
:MEASure:RTIMe:SCURrent
:MEASure:RTIMe:SAVerage
:MEASure:RTIMe:SDEViation