RIGOL 2 Command System
2-12 DS2000E Programming Guide
:ACQuire:MDEPth
Syntax
:ACQuire:MDEPth <mdep>
:ACQuire:MDEPth?
Description
Set or query the memory depth of the oscilloscope; namely the number of waveform points that can be
stored in a single trigger sample. The default unit is pts (points).
Parameter
Name Type Range Default
<mdep> Discrete Refer to Explanation AUTO
Explanation
Non-interweave: AUTO|7000|70000|700000|7000000|28000000 (option).
Return Format
The query returns the memory depth of the oscilloscope in integer.
:ACQuire:SRATe?
Syntax
:ACQuire:SRATe?
Description
Query the current sample rate. The default unit is Sa/s.
Return Format
The query returns the current sample rate in scientific notation. For example, 1.00000e+09; namely
1.000GSa/s.