RIGOL 2 Command System
2-2 DS2000E Programming Guide
:AUToscale
Syntax
:AUToscale
Description
Enable the waveform auto setting function. The oscilloscope adjusts the vertical scale, horizontal time
base and trigger mode according to the input signal automatically to realize optimum waveform display.
The function of this command is the same with that of AUTO on the front panel.
Explanation
This command is invalid when the AUTO function is disabled. For the details, please refer
to
:SYSTem:AUToscale.
When the current status of the pass/fail function is "Enable Test", this command is not available. For
the details, please refer to
:MASK:ENABle.
:CLEar
Syntax
:CLEar
Description
Clear all the waveforms on the screen. New waveforms will still be displayed if the oscilloscope is in RUN
state. The function of this command is the same with that of CLEAR on the front panel.
:RUN
:STOP
Syntax
:RUN
:STOP
Description
The :RUN command starts the oscilloscope and the :STOP command stops the oscilloscope. The functions
of these commands are the same with those of RUN/STOP on the front panel.
Explanation
These commands are invalid when waveform recording or waveform playback is turned on.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals