RIGOL 2 Command System
2-182 DS2000E Programming Guide
:MEASure:R2FPhase
Syntax 1
:MEASure:R2FPhase <chanA>,<chanB>
Description 1
Enable the phase deviation (rising edge-falling edge) measurement function between the two specified
channels.
Syntax 2
:MEASure:R2FPhase? <chanA>,<chanB>
Description 2
Enable the phase deviation (rising edge-falling edge) measurement function between the two specified
channels and return the measurement result (the unit is degree).
Parameter
Name Type Range Default
<chanA> Discrete {CHANnel1|CHANnel2} --
<chanB> Discrete {CHANnel1|CHANnel2} --
Return Format
The query returns the phase deviation (rising edge-falling edge) measurement result in scientific notation.
When the measurement result is invalid, the query returns 9.9e37.
Example
:MEASure:R2FPhase CHANnel1,CHANnel2
:MEASure:R2FPhase? CHANnel1,CHANnel2
You can use the following commands to enable the phase deviation (rising edge-falling edge)
measurement function between the two specified channels and the statistic function of the measurement
results as well as query the statistic maximum value, minimum value, current value, average value, and
standard deviation of the phase deviation (rising edge-falling edge) measurement results.
:MEASure:R2FPhase:SMAXimum
:MEASure:R2FPhase:SMINimum
:MEASure:R2FPhase:SCURrent
:MEASure:R2FPhase:SAVerage
:MEASure:R2FPhase:SDEViation