2 Command System RIGOL
DS2000E Programming Guide
2-159
:MEASure:FTIMe:SMAXimum
:MEASure:FTIMe:SMINimum
:MEASure:FTIMe:SCURrent
:MEASure:FTIMe:SAVerage
:MEASure:FTIMe:SDEViation
Syntax 1
:MEASure:FTIMe:SMAXimum <chan>
:MEASure:FTIMe:SMINimum <chan>
:MEASure:FTIMe:SCURrent <chan>
:MEASure:FTIMe:SAVerage <chan>
:MEASure:FTIMe:SDEViation <chan>
Description 1
Enable the fall time measurement function of the specified channel and the statistic function of the
measurement results.
Syntax 2
:MEASure:FTIMe:SMAXimum? <chan>
:MEASure:FTIMe:SMINimum? <chan>
:MEASure:FTIMe:SCURrent? <chan>
:MEASure:FTIMe:SAVerage? <chan>
:MEASure:FTIMe:SDEViation? <chan>
Description 2
Query the statistic maximum value (SMAXimum), minimum value (SMINimum), current value (SCURrent),
average value (SAVerage), or standard deviation (SDEViation) of the fall time measurement results when
the fall time measurement function of the specified channel and the statistic function of the measurement
results are enabled.
Parameter
Name Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the statistic maximum value, minimum value, current value, average value, or standard
deviation of the fall time measurement results in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals