EasyManuals Logo

Rigol DS2000E Series Programming Guide

Rigol DS2000E Series
365 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #219 background imageLoading...
Page #219 background image
2 Command System RIGOL
DS2000E Programming Guide
2-203
:MEASure:MARea:SMAXimum
:MEASure:MARea:SMINimum
:MEASure:MARea:SCURrent
:MEASure:MARea:SAVerage
:MEASure:MARea:SDEViation
Syntax 1
:MEASure:MARea:SMAXimum <chan>
:MEASure:MARea:SMINimum <chan>
:MEASure:MARea:SCURrent <chan>
:MEASure:MARea:SAVerage <chan>
:MEASure:MARea:SDEViation <chan>
Description 1
Enable the area measurement function of the specified channel and the statistic function of the
measurement results.
Syntax 2
:MEASure:MARea:SMAXimum? <chan>
:MEASure:MARea:SMINimum? <chan>
:MEASure:MARea:SCURrent? <chan>
:MEASure:MARea:SAVerage? <chan>
:MEASure:MARea:SDEViation? <chan>
Description 2
Query the statistic maximum value (SMAXimum), minimum value (SMINimum), current value (SCURrent),
average value (SAVerage), or standard deviation (SDEViation) of the area measurement results when the
area measurement function of the specified channel and the statistic function of the measurement results
are enabled.
Parameter
Name
Type Range Default
<chan> Discrete {CHANnel1|CHANnel2|MATH} --
Return Format
The query returns the statistic maximum value, minimum value, current value, average value, or standard
deviation of the area measurementresults in scientific notation. When the measurement results are
invalid, the query returns 9.9e37.

Table of Contents

Other manuals for Rigol DS2000E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS2000E Series and is the answer not in the manual?

Rigol DS2000E Series Specifications

General IconGeneral
BrandRigol
ModelDS2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals