8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
148 List of line codes for extended HIF
q oIf no calibrator lot number was defined, "??????????" will
be used and transmitted (without quotes).
o If no calibrator lot expiration date is given,
"00/00/0000" will be used and transmitted (without
quotes).
o If no cassette lot number and serial number used for
calibration was defined, "??????????" will be used and
transmitted (without quotes).
o If no ISE Solution 1-3 lot number are given,
"??????????" will be used and transmitted (without
quotes).
o If no ISE Solution 1-3 lot expiration date are given,
"00/00/0000" will be used and transmitted (without
quotes).
o If no Calibration Result Coefficients 1-5 are given,
"??????????" will be used and transmitted (without
quotes).
o If the any lot number or coefficient labels/values are
shorter than ten characters, the string will be filled
with spaces on the right side.
Example, 68354 will be 68354 ••••••
(•) represents a filler space
o If the calibrator flag is shorter than three characters,
the string will be filled with spaces on the left side.
Example, 0 will be••0
(•) represents a filler space
Flag code Description
0 No Flag
1 Not enough readings in kinetic range for calc mode Kinetic or Kinsearch. (F3)
2 Low Activity. The absorbance change during measurement doesn't reached the accepted limit, i.e . the
reaction has taken off too slow.
3 Absorbance values of the lowest concentration calibrator fall below the lower range. For general
chemistries this is often the reagent blank. (F3)
4 The result is lower than the measurable test range. (F3)
5 Curve Direction is not corresponding with the testdefinition curve direction. (F3)
6 Absorbance values of the lowest concentration calibrator fall is above the upper range. (F3)
7 Deviation between replicates exceed programmed limit. One or more replicates are erroneous. (F3)
8 Recovered values of standards deviate greater than programmed limit. (F3)
9 The result exceeds than the upper measurable test range. (F3)
10 (Serum proteins only) The sample contains an excess of antigen. %1% = [value] - is Test Range High
(F3)
11 Insufficient amount of sample/QC/Solution or too many air bubbles in electrode path. The sample
segment is too short or air bubbles are present, mixing is too vigorous (in the case of the indirect/urine
mode) or a tube is leaking or blocked. (F3)
12 Insufficient calibrator or too many air bubbles. The calibrator segment is too short or air bubbles are
present, or a tube is leaking or blocked. (F3)
15 Nonlinear standard curve cannot be calculated due to calculation error. Action: Check calibration
graph.Rerun calibration with fresh calibrators. (F3)
y Flag codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals