8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
148 List of line codes for extended HIF
q oIf no calibrator lot number was defined, "??????????" will
be used and transmitted (without quotes).
o If no calibrator lot expiration date is given,
"00/00/0000" will be used and transmitted (without
quotes).
o If no cassette lot number and serial number used for
calibration was defined, "??????????" will be used and
transmitted (without quotes).
o If no ISE Solution 1-3 lot number are given,
"??????????" will be used and transmitted (without
quotes).
o If no ISE Solution 1-3 lot expiration date are given,
"00/00/0000" will be used and transmitted (without
quotes).
o If no Calibration Result Coefficients 1-5 are given,
"??????????" will be used and transmitted (without
quotes).
o If the any lot number or coefficient labels/values are
shorter than ten characters, the string will be filled
with spaces on the right side.
Example, 68354 will be 68354 ••••••
(•) represents a filler space
o If the calibrator flag is shorter than three characters,
the string will be filled with spaces on the left side.
Example, 0 will be••0
(•) represents a filler space
Flag code Description
0 No Flag
1 Not enough readings in kinetic range for calc mode Kinetic or Kinsearch. (F3)
2 Low Activity. The absorbance change during measurement doesn't reached the accepted limit, i.e . the
reaction has taken off too slow.
3 Absorbance values of the lowest concentration calibrator fall below the lower range. For general
chemistries this is often the reagent blank. (F3)
4 The result is lower than the measurable test range. (F3)
5 Curve Direction is not corresponding with the testdefinition curve direction. (F3)
6 Absorbance values of the lowest concentration calibrator fall is above the upper range. (F3)
7 Deviation between replicates exceed programmed limit. One or more replicates are erroneous. (F3)
8 Recovered values of standards deviate greater than programmed limit. (F3)
9 The result exceeds than the upper measurable test range. (F3)
10 (Serum proteins only) The sample contains an excess of antigen. %1% = [value] - is Test Range High
(F3)
11 Insufficient amount of sample/QC/Solution or too many air bubbles in electrode path. The sample
segment is too short or air bubbles are present, mixing is too vigorous (in the case of the indirect/urine
mode) or a tube is leaking or blocked. (F3)
12 Insufficient calibrator or too many air bubbles. The calibrator segment is too short or air bubbles are
present, or a tube is leaking or blocked. (F3)
15 Nonlinear standard curve cannot be calculated due to calculation error. Action: Check calibration
graph.Rerun calibration with fresh calibrators. (F3)
y Flag codes