7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface 125
Block 07: Sample Results with Lot Information
Block 07 is transmitted to the host if all of the following
conditions are met:
After receiving a result request with result type
selector 11 or 17
If an accepted and not yet transmitted sample result is
available
Line code Description Iteration
(1)(2)(3)
(1) w = 1 to 141. The value of w can be different from 1 only if the result
was measured using an absorbance test (Line 06). There are 141
possible lines since each line is produced for every cycle (max 141).
(2) p = 0 to x. The value of x is the number of controls used for a single
test result. It is possible that no control was used because QC was
never performed.
(3) y = 1 to z. The value of z is the total number of reruns executed for a
single test result.
53 Order ID
56 Sample Name 0 to 1
15 Access Date Time
16 CALidx
55 Test ID
11 Reagent Lot Information
00 Result data
14 Validation Status
06/08/09 Raw Data 1 to w
17 Test Execution
18 Cassette Information
19 Calibration Data
21 Control Data 0 to p
27 Rerun results 0 to y
y Block 07

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals