5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes 95
Line code 51: Patient Information
Example
51_07/12/1965_M_Ronald•Reagen•1234•on•the•move!_12
34•New•York••••••••<LF>
q The database set-up function allows the user to
disable the patient information fields. Nevertheless, on
the host interface, the birthday, sex and leadertext 1 fields
have to be transmitted even if the values are unknown.
q The entry for patient leadertext 1 appears in the
database as patient name. As the field-identifier Name
cannot be changed and is mandatory for transmission
anyway, this field should only be used to transmit the
patient's name.
Line code 52: Special Order Selection
Example
52_03_T<LF>
Description Format Range
Compulsory Line code 51
Birthday D1 01/01/1800 to 31/12/210000/00/0000 if unknown
Sex S1 M, FX if unknown
Patient leadertext 1 S31 <space>s if unknown
Optional Patient leadertext 2 S21
Patient leadertext 3 S21
y Line code 51: Patient Information
Description Format Range Range description
Line code 52
Special order type I2 01 Calibration order
02 Calibration correction order
03 Control order
Attribute S1 I Immediate
y Line code 52: Special Order Selection

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals