7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
134 Sample order status blocks
Sample order status blocks
In this section
Block 80: Order Status Request (134)
Block 81: Order Status Response (134)
Block 85: Failed Order Status Request (135)
Block 80: Order Status Request
It is possible to retrieve the status of all pending sample
orders or of a specific sample order.
If Line 20 is omitted all the pending sample orders are
retrieved.
If Line 20 is used the status of the specified sample will
be retrieved regardless whether the sample is pending or
not.
Block 81: Order Status Response
This response provides the sample order status for the
requested items.
If the request was performed using block 80 with no line
the total number of pending orders will be returned.
Otherwise if line 20 was used in block 80 only the
requestd sample will be returned (m=1).
Line code Description Iteration
20 Order ID 0 to 1
y Block 80
Line code Description Iteration
(1)(2)
(1) m is the total number of pending sample orders.
(2) n is the total number of blocked test for each sample order.
22 Order ID
1 to m23 Status
25 Test Status 0 to n
y Block 76

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals