8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
156 List of line codes for extended HIF
Line Code 27: Rerun Result
Example
27_•8.694475E+01_•2_1:1•••<LF>
q If the result could not be calculated, the
concentration value is returned as 9.99E+99.
Line Code 28: Test Status
Example
28_17_•0_•••0_N_2_<LF>
Description Format Range Range description
Line Code 27
Sample Concentration F13 +/-9.999999E+99 to +/-1.0E-99
Rerun Order I2
Dilution Flag S6 1:X with X going from 1 to 100
y Line Code 27: Rerun Result
Description Format Range Range description
Line Code 28
Test status I2 3 Blocked
16 Disabled
17 OK
31 Calibration postponed
Blocked status I2 For detailed information, refer to table in line 61.
Control Interval I4 Number of tests or hours between required control.
Refer to next field for unit.
Control Type S1 N none (only if no control interval is set)
H hours
T tests
Calibration due status I1 1 None
2 Next Cassette
3 Next Lot
4 Required
Optional If the calibration is required
Calibration date D1 When is the next calibration due.
y Line Code 28: Test Status