8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface 141
About line codes
Format code Description
Sx x-character ASCII-string, left adjusted.
Length x is fixed.
Shorter strings have to be filled up with trailing <spaces>.
Example: S15 "Mr.•Evans••••••"
Ix ASCII-formatted x-digit integer value, right adjusted.
Length x is fixed.
Shorter integer values have to be filled up with leading <spaces> or zeros (0).
Example: I5 "00069" or "•••69"
SRx.y ASCII-formatted real value with sign, right adjusted.
x = Length of the whole data record.
y = Number of digits after the decimal point (optional).
Length x is fixed.
Shorter real-values have to be filled up with leading <spaces> or zeros (0).
The decimal point is mandatory.
Example: SR9.3: "-0077.123"
Fx ASCII formatted floating point number, right adjusted.
x = Length of the whole data record.
Length x is fixed.
Shorter real-values have to be filled up with leading <spaces> or zeros (0).
The sign "+" or "-" for the exponent is mandatory.
Decimal point and exponent "E" are mandatory.
Mantissa has exactly one digit before the decimal point in the range 1 to 9.
No <spaces> are allowed between the mantissa and the exponent.
Example: F12: "••••6.11E+33" or F15: "-00007.4321E-02"
D1 Date format 1: DD/MM/YYYY.
10 character ASCII-string.
Example: "04/01/1994"
D2 Date format 2: DD-MMM-YY (used by COBAS Core/Core II).
9 character ASCII-string.
Example: "04-JAN-94"
T1 Time format 1: hh:mm:ss (hh in 24 h format).
8 character ASCII-string.
Example: "22:07:34"
T2 Time format 2: hh:mm (hh in 24 h format) (used by COBAS Core/Core II).
5 character ASCII-string.
Example: "22:07"
y Formats of data fields used in line codes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche COBAS INTEGRA 400 plus and is the answer not in the manual?

Roche COBAS INTEGRA 400 plus Specifications

General IconGeneral
BrandRoche
ModelCOBAS INTEGRA 400 plus
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals