7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
130 Calibration status blocks
Calibration status blocks
In this section
Block 52: Calibration status request (130)
Block 54: Calibration Status Response (130)
Block 57: Failed Calibration Status Request (130)
Block 52: Calibration status request
Block 52 is used to interrogate the calibration status.
Block 52 has no lines.
Block 54: Calibration Status Response
This request provides the calibration status for all loaded
tests.
Line 49 is added only if rack position information is
available.
Block 57: Failed Calibration Status Request
Block 57 contains Line 96 and provides the eventual error
encountered following request with Block 52.
Line code Description Iteration
(1)
(1) m is the total number of loaded tests.
55 Test ID
1 to m
60 Calibration due
61 Test Status and
blocked status
62 Calibrator ID 0 to 1
49 Rack Position 0 to 1
y Block 54
Line code Description
96 Error Message
y Block 57