8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
142 List of line codes for extended HIF
List of line codes for extended HIF
Line Code 06: Absorbance Raw Data
Example
06_•49_1_•0.0363_•0.0046<LF>
n is the total number of wavelength used and can go up
to 12. Hence there can be up to 12 floating point numbers
in the same Line.
Line code In Line code In
06 Absorbance Raw Data 36 Cassette definition
08 FP Raw Data 37 Tests left
09 ISE Raw Data 38 Diluent status
14 Validation Status 39 Cassette Status
17 Test Execution 40 Service Selection
18 Cassette Information 45 Service Action Definition
19 Calibration Information 46 Currently Time or Tests elapsed
20 Order ID (Request) 47 Service Due
21 Control Information 48 Elapsed percentage
22 Order ID (Response) 49 Rack Position
23 Sample Order Status 55 Test ID
24 Message Date Time 58 Service action result
25 Test Order Status 60 Calibration due
26 Dilution Factor 61 Test Status and blocked status
27 Rerun Result 62 Calibrator ID
28 Test Status 82 Message Icon
29 Cassette status 83 Message String
30 Operation selection 93 System Date Time
31 Calibrator status 94 Serial Number
32 Service Action selection 97 System status
33 Control status
y List of line codes
Description Format Range
Line Code 06
Cycle I3 1 to 141
Replicate Number I1 1 to 3
Absorbance Data F7 1 to n
y Line Code 06: Absorbance Raw Data